Set the systolic correction by pressing and hold- ing the CALIBRATE key to the right of the GAP key. Using the arrow keys, set the correction. For example, if the blood pressure reading for systolic pressure was 148 mmHg, the systolic correction would be +2 and the up arrow key would be pressed until +2 would display in the window.Press the MENU key from the Systolic Correction window to change to the Diastolic Correction window. Using the arrow keys, set the correction. For example, if the blood pressure reading for dia- stolic pressure was 72 mmHg, the diastolic cor-rection would be -2 and the down arrow would be pressed until -2 would display in the window.Press the MENU key. The “CALIBRATION COMPLETE” message will appear and the main menu window will be displayed.