C.3 Macroparticle count C.3.1 Microscopic measurement of particles col dịch - C.3 Macroparticle count C.3.1 Microscopic measurement of particles col Việt làm thế nào để nói

C.3 Macroparticle count C.3.1 Micro

C.3 Macroparticle count

C.3.1 Microscopic measurement of particles collected on filter paper,
see ASTM F312[4]

C.3.2 Cascade impactor, particle collection system where sample is passed at a constant flow rate through
a series of orifices of decreasing dimensions; the orifices face the collection surfaces. As fluid velocity
increases through each orifice-collector stage, smaller particles are collected for weighing or counting after
collection.
The specifications for the cascade impactor are given in Table C.5.

Table C.5 — Specifications for cascade impactor

Iteam Specification
Measuring limits/range Sampling flow rate as specified
Sensitivity/resolution Sub-micrometer particles can be collected at low pressure
Accuracy Stage “cut-point” accuracy is ≥90 %
Linearity Significant quantity of over-and under-size deposition
Stability 50 %. Cutoff size depends on the sample flow rate
Response time Minutes to days, depending on sample measurement method
Calibration interval 12 months maximum

C.3.3 Discrete-macroparticle counter, instrument capable of counting and sizing (when required) single airborne macroparticles.
The specifications for the discret-macroparticle counter are given in Table C.6.

Table C.6 — Specifications for discrete-macroparticle counter

Iteam Specification
Measuring limits/range Particle concentration to 1,0 ×106/m3
Sensitivity/resolution 5 μm to 80 μm with 20 % resolution
Uncertainty of measurement Sizing error ±5 % of calibration setting
Linearity Can vary with particle composition or shape
Calibration interval 12 months maximum
Counting efficiency (50 ±20) % at minimum size threshold and (100 ±10) % for particles greater than or equal to 1,5 times the minimum threshold size.

C.3.4 Time-of-flight particle sizing apparatus, discrete-particle counting and sizing apparatus that defines the aerodynamic diameter of particles by measuring the time for a particle to accommodate to a change in air velocity. This is usually done by measuring the particle transit time optically after a fluid stream velocity
change.
The specifications for the time-of-flight particle sizing apparatus are given in Table C.7.

Table C.7 — Specifications for time-of-flight particle sizing apparatus

Iteam Specification
Measuring limits/range Particle concentration to 1,0 ×107/m3
Sensitivity/resolution 0,5 μm to 20 μm with 20 % resolution
Uncertainty of measurement ±5 % of calibration size setting
Calibration interval 12 months maximum
Counting efficiency (50 ±20) % at minimum size threshold and (100 ±10) % for particles greater than or equal to 1,5 times the minimum threshold size

C.3.5 Piezo-balance impactor, particle collection system where the sample is passed at a constant rate through a series of orifices with decreasing dimensions; the orifices face collection surfaces fitted with piezo-electric quartz microbalance mass sensors which weigh the particle collected by each stage during collection.
The specifications for the piezo-balance impactor are given in Table C.8.

Table C.8 — Specifications for piezo-balance impactor

Iteam Specification
Sensitivity/resolution 5 μm to 50 μm particles collected at low pressures
Linearity Significant quantity of over-and under-size deposition
Stability Cutoff size depends on the sample flow rate
Calibration interval 12 months maximum
Minimum collection sensitivity 10 μg/m3 for particles with specific gravity 2

C.4 Airflow test

C.4.1 Air velocity meter

C.4.1.1 Thermal anemometer, measures air velocity by sensing the change in heat transfer from a small, electrically heated sensor exposed to the airflow.
The specifications for the thermal anemometer are given in Table C.9.


Table C.9 — Specifications for thermal anemometer

Iteam Specification
Measuring limits/range 0,1 m/s to 1,0 m/s typically in the installation, 0,5 m/s to 20 m/s in duct
Sensitivity/resolution 0,05 m/s (or minimum 1 % for full scale)a
Uncertainty of measurement ±(5 % of reading +0,1 m/s)a
Response time
0/5000
Từ: -
Sang: -
Kết quả (Việt) 1: [Sao chép]
Sao chép!
C.3 Macroparticle tính C.3.1 vi đo lường của hạt được thu thập trên giấy lọc, Xem ASTM F312 [4]C.3.2 thác impactor, nơi các mẫu được thông qua tốc độ dòng chảy liên tục thông qua hệ thống hạt một loạt các orifices của giảm kích thước; Các orifices đối mặt với các bề mặt bộ sưu tập. Là vận tốc chất lỏng tăng thông qua từng giai đoạn thu lỗ, hạt nhỏ hơn được thu thập cho nặng hoặc đếm sau khi bộ sưu tập. Thông số kỹ thuật cho cascade impactor được đưa ra trong bảng C.5. Bảng C.5-Thông số kỹ thuật cho thác impactorĐặc điểm kỹ thuật iteamĐo giới hạn/phạm vi tốc độ dòng chảy lấy mẫu theo quy địnhNhạy cảm/độ phân giải tiểu PANME hạt có thể được thu thập ở áp suất thấp Độ chính xác giai đoạn "cắt-điểm" chính xác là ≥90% Số lượng đáng kể linearity của trên- và lắng đọng dưới kích thướcỔn định 50%. Cắt kích thước phụ thuộc vào mẫu lưu lượngPhản ứng thời gian phút đến ngày, tùy thuộc vào phương pháp đo lường mẫuHiệu chuẩn khoảng thời gian 12 tháng tối đaTruy cập Discrete-macroparticle C.3.3, nhạc cụ có khả năng đếm và macroparticles trên không duy nhất định cỡ (khi có yêu cầu). Thông số kỹ thuật số lượt truy cập discret-macroparticle được đưa ra trong bảng C.6. Bảng C.6-Thông số kỹ thuật cho rời rạc macroparticle số lượt truy cậpĐặc điểm kỹ thuật iteamGiới hạn/phạm vi đo nồng độ hạt 1,0 × 106/m3Nhạy cảm/độ phân giải 5 μm đến 80 μm với độ phân giải 20%Không chắc chắn của đo lường định cỡ lỗi ±5% của hiệu chuẩn thiết lập Linearity có thể khác nhau với thành phần hạt hoặc hình dạngHiệu chuẩn khoảng thời gian 12 tháng tối đaTính hiệu quả (50 ±20) % kích thước tối thiểu ngưỡng và % (100 mức ±10) đối với các hạt lớn hơn hoặc bằng với 1,5 lần kích thước tối thiểu ngưỡng.C.3.4 thời gian chuyến bay hạt định cỡ bộ máy, rời rạc hạt đếm và định cỡ bộ máy xác định đường kính khí động học của hạt bằng cách đo thời gian cho một hạt để chứa đến một sự thay đổi trong vận tốc máy. Điều này thường được thực hiện bằng cách đo thời gian vận chuyển hạt quang học sau khi một chất lỏng dòng vận tốc thay đổi. Thông số kỹ thuật cho các hạt thời gian chuyến bay định cỡ bộ máy được đưa ra trong bảng C7. Bảng C7-Thông số kỹ thuật cho thời gian của chuyến bay hạt định cỡ bộ máyĐặc điểm kỹ thuật iteamGiới hạn/phạm vi đo nồng độ hạt 1,0 × 107/m3Nhạy cảm/độ phân giải 0,5 μm đến 20 μm với độ phân giải 20%Không chắc chắn của đo lường ±5% của hiệu chuẩn kích thước thiết lập Hiệu chuẩn khoảng thời gian 12 tháng tối đaTính hiệu quả (50 ±20) % kích thước tối thiểu ngưỡng và % (100 mức ±10) đối với các hạt lớn hơn hoặc bằng với 1,5 lần kích thước tối thiểu ngưỡngC.3.5 số dư Piezo-impactor, Hệ thống thu gom hạt nơi mẫu được thông qua tại một tốc độ không đổi thông qua một loạt các orifices với giảm kích thước; Các orifices đối mặt với bề mặt bộ sưu tập được trang bị với thạch anh piezo-điện microbalance khối lượng bộ cảm biến đó cân nhắc các hạt được thu thập bởi từng giai đoạn trong bộ sưu tập. Thông số kỹ thuật cho impactor piezo-cân bằng được đưa ra trong bảng C.8. Bảng C.8-Chi tiết kỹ thuật áp lực piezo-cân bằng impactorĐặc điểm kỹ thuật iteamNhạy cảm/độ phân giải 5 μm để 50 μm hạt được thu thập ở áp suất thấpSố lượng đáng kể linearity của trên- và lắng đọng dưới kích thướcỔn định Cutoff kích thước phụ thuộc vào mẫu lưu lượngHiệu chuẩn khoảng thời gian 12 tháng tối đaBộ sưu tập tối thiểu nhạy cảm 10 μg/m3 đối với các hạt với tỷ trọng riêng 2 C.4 khí thử nghiệm C.4.1 máy vận tốc mét C.4.1.1 biết nhiệt, các biện pháp máy vận tốc bởi cảm biến sự thay đổi trong truyền nhiệt từ một cảm biến nhỏ, điện nước nóng tiếp xúc với luồng không khí. Thông số kỹ thuật cho biết nhiệt được đưa ra trong bảng C.9. Bảng C.9-Thông số kỹ thuật cho biết nhiệtĐặc điểm kỹ thuật iteamĐo giới hạn/nhiều 0,1 m/s để 1,0 m/s thường trong tiến trình cài đặt, 0,5 m/s đến 20 m/s trong ốngNhạy cảm/độ phân giải 0,05 m/s (hoặc tối thiểu 1% cho quy mô đầy đủ) mộtKhông chắc chắn của đo lường ± (5% của đọc + 0,1 m/s) mộtThời gian đáp ứng < 1 s 90% của quy mô đầy đủ Hiệu chuẩn khoảng thời gian 12 tháng tối đamột cho nhạy cảm và không chắc chắn của đo lường, đề cập đến ISO 7726. Bộ máy cần chỉnh để không khí khác biệt nhiệt độ và thay đổi của áp suất khí quyển.C.4.1.2 siêu âm biết, chiều 3 hoặc tương đương, các biện pháp máy vận tốc bởi cảm biến sự chuyển đổi của tần số âm thanh (hoặc vận tốc âm thanh) tách ra điểm trong luồng không khí đo. The specifications for the ultrasonic anemometer are given in Table C.10. Table C.10 — Specifications for ultrasonic anemometer, 3-dimensional or equivalentIteam SpecificationMeasuring limits/range 0 m/s to 1 m/s in the installationSensitivity/resolution 0,01 m/sUncertainty of measurement ±5 % of reading Response time <1 sCalibration interval 12 months maximumC.4.1.3 Vane-type anemometer, measures air velocity by counting the revolution rate of the anemometer vanes the airflow. The specifications for the vane-type anemometer are given in Table C.11. Table C.11 — Specifications for vane-type anemometer Iteam SpecificationMeasuring limits/range 0,2 m/s to 10 m/sSensitivity/resolution 0,1 m/sUncertainty of measurement ±0,2 m/s or ±5 % of reading, whichever is greater Response time <10 s at 90 % of full scale Calibration interval 12 months maximumC.4.1.4 Pitot-static tubes and manometer (digital), measure air velocity from the difference of total and static pressures at a position in the airflow, using electrical digital manometers. The specifications for the Pitot-static tube and manometer are given in Table C.12.Table C.12 — Specifications for Pitot-static tube and manometerIteam SpecificationMeasuring limits/range >1,5 m/s Sensitivity/resolution 0,5 m/sUncertainty of measurement ±5 % of reading Response time <10 s at 90 % of full scale Calibration interval 12 months maximumC.4.2 Airflow meter C.4.2.1 Flowhood with flowmeter, measures airflow rate from an area over which there can be variations in airflow, providing an integrated air volume from that area. The total airflow is collected and concentrated so that the velocity at the measurement point represents the cross-sectional average velocity from the total area. The specifications for the flowhood with flowmeter are given in Table C.13. Table C.13 — Specifications for flowhood with flowmeterIteam SpecificationMeasuring limits/range Flow rate of 503m/hr to at least 1 700 m3/hraUncertainty of measurement ±5 % of readingResponse time <10 s at 90 % Calibration interval 12 months maximuma Typical range for size 600 ×600 mm hood. Measuring limits and the resolution depend on the size of hood used.C.4.2.2 Orifice meter, refer to ISO 5167-2:2003[20]. C.4.2.3 Venturi meter, refer to ISO 5167-4:2003[22]. C.5 Air pressure difference test C.5.1 Electronic micromanometer, used to display or output the value of the air pressure difference between a space and its surroundings by detecting the change of electrostatic capacitance or electronic resistance due to the displacement of a diaphragm. The specifications for the electronic micromanometer are given in Table C.14. Table C.14 — Specifications for electronic micromanometerIteam SpecificationMeasuring limits/range 0 kPa to 0,3 kPa, or 0 kPa to 1,5 kPaSensitivity 1 Pa for 0 kPa to 0,3 kPaUncertainty of measurement ±3 % for 0 kPa to 0,3 kPaScale amplitude power 2 (at minimum) to 10 at 0 kPa to 0,3 kPaC.5.3 Mechanical differential pressure gauge, used to measure the air pressure difference between two areas by detecting the movement distance of a needle connected with a mechanical gear or magnetic linkage to the displacement of a diaphragm. The specifications for the mechanical differential pressure gauge are given in Table C.16. Table C.16 — Specifications for mechanical differential pressure gaugeIteam SpecificationMeasuring limits/range 0 Pa to 50 Pa for a small range; 0 kPa to 50 kPa for a large rangeSensitivity/resolution 0,5 Pa for 0 Pa to 50 Pa rangeUncertainty of measurement ±5 % full-scale for 0 Pa to 50 Pa ±2,5 % full-scale for 0 kPa to 50 kPa C.6 Installed filter system leakage test C.6.1 Aerosol photometers C.6.1.1 Linear aerosol photometer, used to measure the mass concentration of aerosols in micrograms per litre (μg/l). The photometer uses a forward scattered-light optical chamber to make this measurement. This apparatus may be used to measure filter leak penetration directly. The specifications for the linear aerosol photometer are given in Table C.17. Table C.17 — Specifications for linear aerosol photometerIteam SpecificationMeasuring limits/range 0,001 μg/l to 100 μg/l −5 full linear decades Sensitivity/resolution 0,001 μg/lUncertainty of measurement ±5 %
Linearity ±0,5 %
Stability ±0,002 μg/l per minute
Response time From 0 % to 90 %, ≤30 s; from 100 μg/l to 10 g/l, ≤60 s
Calibration interval 12 months or 400 operating hours, whichever is sooner
Sample probe tube length Maximum length is 4 m
Particle size 0,1 μm to 0,6 μm over measuring range
Sample flow Nominal flow rate ±15 %
Sample probe See B.6.2.4
C.6.1.2 Logarithmic aerosol photometer, used to measure the mass concentration of aerosols in micrograms per litre (μg/l). The photometer uses a forward scattered optical chamber to make this measurement. Filter leakage penetration cannot be measured directly on this apparatus.
The specifications for the logarithmic aerosol photometer are given in Table C.18.

Table C.18 — Specifications for logarithmic aerosol photometer

Iteam Specification
Measuring limits/range 0,001 μg/l to 100 μg/l −on one range
Sensitivity/resolution 0,001 μg/l
Uncertainty of measurement ±5 %
Stability ±0,002 μg/l per minute
Response time From 0 % to 90 %, ≤60 s; from 100 μg/l to 10 g/l, ≤90 s
Calibration interval 12 months or 400 operating hours, whichever is sooner
Sample probe tube length Maximum length is 4 m
Particle size 0,1 μ
đang được dịch, vui lòng đợi..
Kết quả (Việt) 2:[Sao chép]
Sao chép!
C.3 Macroparticle đếm C.3.1 hiển vi đo lường của các hạt được thu thập trên giấy lọc, xem ASTM F312 [4] C.3.2 Cascade va chạm, bộ sưu tập hạt hệ thống trong đó mẫu được truyền với tốc độ dòng chảy liên tục qua một loạt các lỗ giảm dần kích thước; sự xuyên phải đối mặt với các bề mặt bộ sưu tập. Là chất lỏng vận tốc tăng lên qua từng giai đoạn lỗ-thu, các hạt nhỏ được thu thập cho cân, đếm sau khi bộ sưu tập. Các thông số kỹ thuật cho các va chạm thác được đưa ra trong Bảng C.5. Bảng C.5 - Thông số kỹ thuật cho cascade va chạm Iteam Đặc điểm kỹ thuật đo giới hạn / tốc độ dòng chảy phạm vi lấy mẫu theo quy định độ nhạy / độ phân giải hạt Sub-micromet có thể được thu thập ở áp suất thấp Stage chính xác "cắt-điểm" chính xác là ≥90% Độ tuyến tính số lượng đáng kể trên và dưới kích thước lắng đọng ổn định 50%. Kích thước cắt phụ thuộc vào tốc độ dòng mẫu Thời gian đáp ứng Minutes to ngày, tùy thuộc vào mẫu phương pháp đo khoảng Calibration 12 tháng tối đa C.3.3 Discrete-macroparticle truy cập, công cụ có khả năng đếm và định cỡ (khi cần) macroparticles không khí duy nhất. Các thông số kỹ thuật cho các discret-macroparticle truy cập được cho trong bảng C.6. Bảng C.6 - Thông số kỹ thuật cho truy cập rời rạc macroparticle Iteam Đặc điểm kỹ thuật đo giới hạn / nồng độ hạt phạm vi đến 1,0 × 106 / m3 Độ nhạy / độ phân giải 5 mm đến 80 mm với 20% Độ phân giải Độ không đảm bảo đo lường Sizing lỗi ± 5% của hiệu chuẩn thiết lập tuyến tính có thể thay đổi theo thành phần hạt hoặc hình dạng khoảng Calibration 12 tháng tối đa hiệu quả đếm (50 ± 20)% ở ngưỡng kích thước tối thiểu và (100 ± 10)% đối với các hạt lớn hơn hoặc bằng đến 1,5 lần kích thước ngưỡng tối thiểu. C.3.4 Thời gian bay của máy kích thước hạt, tính rời rạc hạt và kích thước bộ máy mà xác định đường kính khí động học của các hạt bằng cách đo thời gian cho một hạt để chứa một sự thay đổi trong không khí vận tốc. Điều này thường được thực hiện bằng cách đo thời gian vận chuyển hạt quang sau khi một vận tốc dòng chất lỏng thay đổi. Các thông số kỹ thuật cho các thiết bị cỡ hạt thời gian của chuyến bay được cho trong bảng C.7. Bảng C.7 - Thông số kỹ thuật cho thời gian của chuyến bay bộ máy định cỡ hạt Iteam Đặc điểm kỹ thuật đo giới hạn / khoảng nồng độ hạt to 1,0 × 107 / m3 Độ nhạy / độ phân giải 0,5 mm đến 20 mm với 20% độ phân giải không đảm bảo đo ± 5% của hiệu chỉnh kích thước thiết lập hiệu chuẩn khoảng 12 tháng tối đa hiệu quả đếm (50 ± 20)% ở ngưỡng kích thước tối thiểu và (100 ± 10)% đối với các hạt lớn hơn hoặc bằng 1,5 lần kích thước ngưỡng tối thiểu C.3.5 Piezo-balance va chạm, hệ thống thu gom hạt trong đó mẫu được thông qua tại tốc độ không đổi thông qua một loạt các các lỗ với giảm kích thước; sự xuyên phải đối mặt với các bề mặt bộ sưu tập trang bị cảm biến khối thạch anh cân vi áp điện có cân nặng hạt được thu thập bởi mỗi giai đoạn trong quá trình thu. Các thông số kỹ thuật cho các va chạm Piezo-toán được đưa ra trong Bảng C.8. Bảng C.8 - Thông số kỹ thuật cho piezo- cân bằng vật va chạm Iteam Đặc điểm kỹ thuật nhạy / độ phân giải 5 micron đến 50 micron hạt thu thập ở áp suất thấp Độ tuyến tính số lượng đáng kể trên và dưới kích thước lắng đọng kích thước ổn định Cutoff phụ thuộc vào mẫu tốc độ dòng chảy Calibration khoảng thời gian 12 tháng tối đa độ nhạy thu tối thiểu 10 mg / m3 cho hạt có trọng lượng riêng 2 C.4 Airflow test C.4.1 Air vận tốc đo C.4.1.1 máy đo gió nhiệt, đo vận tốc không khí bằng cách cảm nhận sự thay đổi trong chuyển nhiệt từ một, cảm biến nhiệt bằng điện nhỏ tiếp xúc với không khí. Các thông số kỹ thuật cho các máy đo gió nhiệt được đưa ra trong Bảng C.9. Bảng C.9 - Thông số kỹ thuật cho máy đo gió nhiệt Iteam Đặc điểm kỹ thuật đo giới hạn / phạm vi 0,1 m / s đến 1,0 m / s thường trong việc cài đặt, 0,5 m / s đến 20 m / s trong ống nhạy / độ phân giải 0,05 m / s (hoặc tối thiểu 1% đối với quy mô đầy đủ) là một sự không chắc chắn đo ± (5% đọc +0,1 m / s) một thời gian đáp ứng <1 s 90 % của toàn thang khoảng Calibration 12 tháng tối đa một Đối với độ nhạy và độ không đảm bảo đo, hãy tham khảo ISO 7726. Các bộ máy cần tính đến điều khác biệt nhiệt độ không khí và những thay đổi của áp suất khí quyển. C.4.1.2 siêu âm đo gió, 3 chiều hoặc tương đương, đo vận tốc không khí bằng cách cảm nhận sự thay đổi của tần số âm thanh (hay vận tốc âm thanh) giữa các điểm riêng biệt trong luồng không khí đo được. Các thông số kỹ thuật cho các máy đo gió siêu âm được đưa ra trong Bảng C.10. Bảng C.10 - Thông số kỹ thuật cho máy đo gió siêu âm, 3 chiều hoặc tương đương Iteam Đặc điểm kỹ thuật đo giới hạn / range 0 m / s đến 1 m / s trong các cài đặt độ nhạy / độ phân giải 0,01 m / s Độ không đảm bảo đo ± 5% giá trị đọc Thời gian đáp ứng <1 s Calibration khoảng thời gian 12 tháng tối đa C .4.1.3 cánh quạt máy đo gió, đo vận tốc không khí bằng cách đếm tốc độ vòng quay của máy đo gió cánh hướng dòng không khí. Các thông số kỹ thuật cho các loại máy đo gió cánh được cho trong bảng C.11. Bảng C.11 - Thông số kỹ thuật cho cánh-type máy đo gió Iteam Đặc điểm kỹ thuật đo giới hạn / phạm vi 0,2 m / s đến 10 m / s Độ nhạy / độ phân giải 0,1 m / s Độ không đảm bảo đo ± 0,2 m / s hoặc ± 5% giá trị đọc, nào là lớn hơn thời gian đáp ứng < 10 s ở 90% của toàn thang khoảng Calibration 12 tháng tối đa ống Pitot C.4.1.4 tĩnh và áp kế (kỹ thuật số), đo vận tốc không khí từ sự chênh lệch áp lực tổng và tĩnh tại một vị trí trong luồng không khí, sử dụng áp kế kỹ thuật số điện . Các thông số kỹ thuật cho các ống Pitot tĩnh và áp kế được đưa ra trong Bảng C.12. Bảng C.12 - Thông số kỹ thuật cho ống Pitot tĩnh và áp kế Iteam Đặc điểm kỹ thuật đo giới hạn / range> 1,5 m / s Độ nhạy / độ phân giải 0, 5 m / s Độ không đảm bảo đo ± 5% giá trị đọc Thời gian đáp ứng <10 s ở 90% của toàn thang khoảng Calibration 12 tháng tối đa C.4.2 Airflow meter C.4.2.1 Flowhood với lưu lượng kế, đo tốc độ luồng khí từ một khu vực trên đó có thể có các biến trong luồng không khí, cung cấp một khối lượng không khí tích hợp từ khu vực đó. Tổng luồng khí được thu thập và tập trung để các vận tốc tại điểm đo đại diện cho vận tốc trung bình cắt ngang từ tổng diện tích. Các thông số kỹ thuật cho các flowhood với lưu lượng kế được đưa ra trong Bảng C.13. Bảng C.13 - Thông số kỹ thuật cho flowhood với lưu lượng kế Iteam Đặc điểm kỹ thuật đo giới hạn / phạm vi tỷ lệ dòng chảy của 503m / hr để ít nhất 1 700 m3 / HRA không đảm bảo đo ± 5% giá trị đọc Thời gian đáp ứng <10 s ở 90% khoảng thời gian hiệu chuẩn 12 tháng tối đa một loạt điển hình với kích thước 600 × 600 mm mui xe. Đo giới hạn và độ phân giải phụ thuộc vào kích thước của mui xe được sử dụng. C.4.2.2 Orifice mét, tham khảo ISO 5167-2:. 2003 [20] C.4.2.3 Venturi mét, tham khảo ISO 5167-4: 2003 [ 22] . C.5 Air kiểm tra sự khác biệt áp micromanometer C.5.1 điện tử, được sử dụng để hiển thị hoặc ra giá trị của sự khác biệt áp suất không khí giữa một không gian và môi trường xung quanh của nó bằng cách phát hiện sự thay đổi của điện dung tĩnh điện hoặc kháng điện tử do sự dịch chuyển của một . diaphragm Các thông số kỹ thuật cho các micromanometer điện tử được đưa ra trong Bảng C.14. Bảng C.14 - Thông số kỹ thuật cho micromanometer điện tử Iteam Đặc điểm kỹ thuật đo giới hạn / range 0 kPa đến 0,3 kPa, hoặc 0 kPa đến 1,5 kPa Độ nhạy 1 Pa 0 kPa đến 0,3 kPa Độ không đảm bảo đo ± 3% cho 0 kPa đến 0,3 kPa Scale biên độ điện 2 (tối thiểu) đến 10 0 kPa đến 0,3 kPa C.5.3 Cơ đo áp suất khác biệt, được sử dụng để đo chênh lệch áp suất không khí giữa hai khu vực bằng cách phát hiện khoảng cách di chuyển của một cây kim được kết nối với một thiết bị cơ khí hoặc liên kết từ tính đến việc di tản một màng. Các thông số kỹ thuật cho các máy đo áp suất khác biệt cơ khí được đưa ra trong Bảng C.16. Bảng C. 16 - Thông số kỹ thuật cho máy đo áp suất khác biệt cơ khí Iteam Đặc điểm kỹ thuật đo giới hạn / khoảng 0 đến 50 Pa Pa cho một phạm vi nhỏ; 0 kPa đến 50 kPa cho một phạm vi rộng lớn nhạy / độ phân giải 0,5 Pa 0 đến 50 Pa Pa phạm vi không đảm bảo đo ± 5% quy mô toàn cho 0 đến 50 Pa Pa ± 2,5% quy mô toàn cho 0 kPa đến 50 kPa C.6 lọc cài đặt thử nghiệm hệ thống rò rỉ C.6.1 Aerosol trắc quang C.6.1.1 Tuyến tính aerosol quang, được sử dụng để đo nồng độ khối lượng của hạt nhân ngưng tụ trong microgram trên một lít (mg / l). Các quang sử dụng một buồng quang học chuyển tiếp phân tán ánh sáng để thực hiện phép đo này. Thiết bị này có thể được sử dụng để đo lường lọc thâm nhập rò rỉ trực tiếp. Các thông số kỹ thuật cho các aerosol quang tuyến tính được đưa ra trong Bảng C.17. Bảng C.17 - Thông số kỹ thuật cho aerosol tuyến tính quang Iteam Đặc điểm kỹ thuật đo giới hạn / phạm vi 0,001 mg / l đến 100 mg / l -5 thập kỷ tuyến tính đầy đủ nhạy / độ phân giải 0001 mg / l Độ không đảm bảo đo ± 5% Độ tuyến tính: ± 0,5% Tính ổn định ± 0.002 mg / l mỗi phút Thời gian đáp ứng từ 0% đến 90%, ≤30 s; từ 100 mg / l đến 10 g / l, ≤60 s Calibration khoảng thời gian 12 tháng hoặc 400 giờ hoạt động, nào là sớm thăm dò mẫu chiều dài ống dài tối đa là 4 m Kích thước hạt 0,1 mm đến 0,6 mm trên dải đo mẫu tốc độ dòng chảy dòng chảy danh nghĩa ± 15% mẫu dò See B.6.2.4 C.6.1.2 aerosol Logarithmic quang, được sử dụng để đo nồng độ khối lượng của hạt nhân ngưng tụ trong microgram trên một lít (mg / l). Các quang sử dụng một buồng quang phía trước rải rác để thực hiện đo lường này. Xuyên qua bộ lọc rò rỉ không thể đo trực tiếp trên thiết bị này. Các thông số kỹ thuật cho các aerosol quang logarit được đưa ra trong Bảng C.18. Bảng C.18 - Thông số kỹ thuật cho aerosol logarit quang Iteam Đặc điểm kỹ thuật đo giới hạn / phạm vi 0,001 mg / l đến 100 mg / l -on một phạm vi độ nhạy / độ phân giải 0001 mg / l Độ không đảm bảo đo ± 5% Tính ổn định ± 0.002 mg / l mỗi phút Thời gian đáp ứng từ 0% đến 90%, ≤60 s; từ 100 mg / l đến 10 g / l, ≤90 s Calibration khoảng thời gian 12 tháng hoặc 400 giờ hoạt động, nào là sớm thăm dò mẫu chiều dài ống dài tối đa là 4 m hạt kích thước 0,1 μ
































































































































































































đang được dịch, vui lòng đợi..
 
Các ngôn ngữ khác
Hỗ trợ công cụ dịch thuật: Albania, Amharic, Anh, Armenia, Azerbaijan, Ba Lan, Ba Tư, Bantu, Basque, Belarus, Bengal, Bosnia, Bulgaria, Bồ Đào Nha, Catalan, Cebuano, Chichewa, Corsi, Creole (Haiti), Croatia, Do Thái, Estonia, Filipino, Frisia, Gael Scotland, Galicia, George, Gujarat, Hausa, Hawaii, Hindi, Hmong, Hungary, Hy Lạp, Hà Lan, Hà Lan (Nam Phi), Hàn, Iceland, Igbo, Ireland, Java, Kannada, Kazakh, Khmer, Kinyarwanda, Klingon, Kurd, Kyrgyz, Latinh, Latvia, Litva, Luxembourg, Lào, Macedonia, Malagasy, Malayalam, Malta, Maori, Marathi, Myanmar, Mã Lai, Mông Cổ, Na Uy, Nepal, Nga, Nhật, Odia (Oriya), Pashto, Pháp, Phát hiện ngôn ngữ, Phần Lan, Punjab, Quốc tế ngữ, Rumani, Samoa, Serbia, Sesotho, Shona, Sindhi, Sinhala, Slovak, Slovenia, Somali, Sunda, Swahili, Séc, Tajik, Tamil, Tatar, Telugu, Thái, Thổ Nhĩ Kỳ, Thụy Điển, Tiếng Indonesia, Tiếng Ý, Trung, Trung (Phồn thể), Turkmen, Tây Ban Nha, Ukraina, Urdu, Uyghur, Uzbek, Việt, Xứ Wales, Yiddish, Yoruba, Zulu, Đan Mạch, Đức, Ả Rập, dịch ngôn ngữ.

Copyright ©2025 I Love Translation. All reserved.

E-mail: