1.2 Limit defining parameters1.2.1 Detection limit of the analytical p dịch - 1.2 Limit defining parameters1.2.1 Detection limit of the analytical p Việt làm thế nào để nói

1.2 Limit defining parameters1.2.1

1.2 Limit defining parameters
1.2.1 Detection limit of the analytical procedure

The detection limit of the analytical procedure is 2.60 pg. This is the amount of toluene that will give an instrument response that is significantly different from the background response of a reagent blank. (Sections 4.1 and 4.2)

1.2.2 Detection limit of the overall procedure

The detection limits of the overall procedure are 246 ng per sample (5.4 ppb or 20.5 µg/m³), 344 ng per sample (7.6 ppb or 28.7 µg/m³), 657 ng per sample (25 ppb or 93 µg/m³) and 904 ng per sample (67 ppb or 253 µg/m³) for charcoal tubes, Anasorb® 747 tubes, 3M 3520 OVMs and SKC 575-002 samplers respectively. These are the amounts of toluene spiked on the respective samplers that will give instrument responses that are significantly different from the background responses of respective sampler blanks. (Sections 4.1 and 4.3)

1.2.3 Reliable quantitation limit

The reliable quantitation limits are 820 ng per sample (18.1 ppb or 68.3 µg/m³), 1146 ng per sample (25.4 ppb or 95.5 µg/m³), 2190 ng per sample (82 ppb or 309 µg/m³) and 3012 ng per sample (224 ppb or 844 µg/m³) for charcoal tubes, Anasorb® 747 tubes, 3M 3520 OVMs and SKC 575-002 samplers respectively. These are the amounts of toluene spiked on the respective samplers that will give signals that are considered the lower limits for precise quantitative measurements. (Section 4.4)

1.2.4 Precision (analytical procedure)

The precisions of the analytical procedure, measured as the pooled relative standard deviations from standards over concentration ranges equivalent to 0.5 to 2 times the TWA target concentration, are 0.76%, 0.94% and 0.76% for the adsorbent tubes, 3M 3520 OVMs and SKC 575-002 samplers respectively. (Section 4.5)

1.2.5 Precision (overall procedure
1.2.5.1 Adsorbent tubes samples

The precisions of the overall procedure at the 95% confidence level from the ambient temperature storage tests for TWA, ceiling and peak samples for charcoal tubes and Anasorb® 747 tubes are given in Table 1.2.5.1. The TWA samples are 240-min samples taken from 200-ppm atmospheres, the ceiling samples are 10-min samples taken from 300-ppm atmospheres and the peak samples are 1-min samples taken from 500-ppm atmospheres. (Section 4.6)
0/5000
Từ: -
Sang: -
Kết quả (Việt) 1: [Sao chép]
Sao chép!
1.2 giới hạn xác định tham số1.2.1 phát hiện giới hạn của các thủ tục phân tích Giới hạn phát hiện của các thủ tục phân tích là 2,60 pg. Đây là số lượng toluene sẽ cung cấp cho một phản ứng cụ là đáng kể khác nhau từ các phản ứng nền của một trống tinh khiết. (Phần 4.1 và 4.2) 1.2.2 phát hiện giới hạn của các thủ tục tổng thể Các giới hạn phát hiện của các thủ tục tổng thể là 246 ng mỗi mẫu (5,4 ppb hoặc 20,5 μg/m³), 344 ng mỗi mẫu (7.6 ppb hoặc 28,7 μg/m³), 657 ng mỗi mẫu (25 ppb hoặc 93 μg/m³) và 904 ng mỗi mẫu (67 ppb hoặc 253 μg/m³) cho than ống, ống Anasorb ® 747, 3 M 3520 OVMs và SKC 575-002 samplers tương ứng. Đây là số tiền của toluene tăng vọt ngày các samplers tương ứng sẽ cung cấp cho phản ứng cụ là đáng kể khác nhau từ các phản ứng nền tương ứng sampler trống. (Phần 4.1 và 4.3) 1.2.3 đáng tin cậy quantitation giới hạn Các giới hạn đáng tin cậy quantitation là 820 ng mỗi mẫu (18.1 ppb hoặc 68,3 μg/m³), 1146 ng mỗi mẫu (25,4 ppb hoặc 95,5 μg/m³), 2190 ng mỗi mẫu (82 ppb hoặc 309 μg/m³) và 3012 ng mỗi mẫu (224 ppb hoặc 844 μg/m³) cho than ống, ống Anasorb ® 747, 3 M 3520 OVMs và SKC 575-002 samplers tương ứng. Đây là số tiền của toluene tăng vọt ngày các samplers tương ứng sẽ cung cấp cho tín hiệu được coi là các giới hạn thấp hơn để đo lường định lượng chính xác. (Phần 4.4) 1.2.4 Precision (analytical procedure) The precisions of the analytical procedure, measured as the pooled relative standard deviations from standards over concentration ranges equivalent to 0.5 to 2 times the TWA target concentration, are 0.76%, 0.94% and 0.76% for the adsorbent tubes, 3M 3520 OVMs and SKC 575-002 samplers respectively. (Section 4.5) 1.2.5 Precision (overall procedure1.2.5.1 Adsorbent tubes samples The precisions of the overall procedure at the 95% confidence level from the ambient temperature storage tests for TWA, ceiling and peak samples for charcoal tubes and Anasorb® 747 tubes are given in Table 1.2.5.1. The TWA samples are 240-min samples taken from 200-ppm atmospheres, the ceiling samples are 10-min samples taken from 300-ppm atmospheres and the peak samples are 1-min samples taken from 500-ppm atmospheres. (Section 4.6)
đang được dịch, vui lòng đợi..
Kết quả (Việt) 2:[Sao chép]
Sao chép!
1.2 Giới hạn xác định các thông số
1.2.1 Giới hạn phát hiện của các thủ tục phân tích Giới hạn phát hiện của các thủ tục phân tích là 2.60 pg. Đây là số tiền của toluene rằng sẽ đưa ra một phản ứng cụ đó là khác nhau đáng kể từ các phản ứng của một nền trống thuốc thử. (Mục 4.1 và 4.2) 1.2.2 Giới hạn phát hiện của thủ tục chung Các giới hạn phát hiện của thủ tục tổng thể là 246 ng mỗi mẫu (5,4 ppb hoặc 20,5 mg / m³), 344 ng mỗi mẫu (7,6 ppb hoặc 28,7 mg / m³) , 657 ng mỗi mẫu (25 ppb hoặc 93 mg / m³) và 904 ng mỗi mẫu (67 ppb hoặc 253 mg / m³) cho các ống than, Anasorb® 747 ống, 3M 3520 OVMs và SKC 575-002 lấy mẫu tương ứng. Đây là số tiền của toluene tăng vọt trên lấy mẫu tương ứng mà sẽ cung cấp cho phản ứng cụ mà là khác nhau đáng kể từ các phản ứng nền của khoảng trống sampler tương ứng. (Mục 4.1 và 4.3) 1.2.3 Giới hạn định lượng đáng tin cậy Các giới hạn định lượng đáng tin cậy là 820 ng mỗi mẫu (18,1 ppb hoặc 68,3 mg / m³), 1146 ng mỗi mẫu (25,4 ppb hoặc 95,5 mg / m³), 2190 ng mỗi mẫu ( 82 ppb hoặc 309 mg / m³) và 3012 ng mỗi mẫu (224 ppb hoặc 844 mg / m³) cho các ống than, Anasorb® 747 ống, 3M 3520 OVMs và SKC 575-002 lấy mẫu tương ứng. Đây là số tiền của toluene tăng vọt trên lấy mẫu tương ứng mà sẽ cho tín hiệu được coi là giới hạn thấp hơn cho các phép đo định lượng chính xác. (Phần 4.4) 1.2.4 Precision (thủ tục phân tích) Các độ chính xác của thủ tục phân tích, đo là độ lệch chuẩn tương đối tổng hợp từ các tiêu chuẩn về nồng độ dao động tương đương với 0,5-2 lần so với nồng độ mục tiêu TWA, là 0,76%, 0,94% và 0,76 % cho các ống hấp phụ, 3M 3520 OVMs và SKC 575-002 lấy mẫu tương ứng. (Phần 4.5) 1.2.5 Precision (thủ tục chung 1.2.5.1 ống hấp phụ mẫu The độ chính xác của thủ tục chung ở mức độ tin cậy 95% từ các bài kiểm tra lưu trữ nhiệt độ môi trường xung quanh cho TWA, trần và cao điểm mẫu cho các ống than và Anasorb® 747 ống được đưa ra trong Bảng 1.2.5.1. Các mẫu TWA được mẫu 240-min lấy từ bầu khí quyển 200 ppm, các mẫu trần là mẫu 10-min lấy từ bầu khí quyển 300 ppm và các mẫu cao điểm là mẫu 1-min lấy từ 500- bầu khí quyển ppm. (phần 4.6)



















đang được dịch, vui lòng đợi..
 
Các ngôn ngữ khác
Hỗ trợ công cụ dịch thuật: Albania, Amharic, Anh, Armenia, Azerbaijan, Ba Lan, Ba Tư, Bantu, Basque, Belarus, Bengal, Bosnia, Bulgaria, Bồ Đào Nha, Catalan, Cebuano, Chichewa, Corsi, Creole (Haiti), Croatia, Do Thái, Estonia, Filipino, Frisia, Gael Scotland, Galicia, George, Gujarat, Hausa, Hawaii, Hindi, Hmong, Hungary, Hy Lạp, Hà Lan, Hà Lan (Nam Phi), Hàn, Iceland, Igbo, Ireland, Java, Kannada, Kazakh, Khmer, Kinyarwanda, Klingon, Kurd, Kyrgyz, Latinh, Latvia, Litva, Luxembourg, Lào, Macedonia, Malagasy, Malayalam, Malta, Maori, Marathi, Myanmar, Mã Lai, Mông Cổ, Na Uy, Nepal, Nga, Nhật, Odia (Oriya), Pashto, Pháp, Phát hiện ngôn ngữ, Phần Lan, Punjab, Quốc tế ngữ, Rumani, Samoa, Serbia, Sesotho, Shona, Sindhi, Sinhala, Slovak, Slovenia, Somali, Sunda, Swahili, Séc, Tajik, Tamil, Tatar, Telugu, Thái, Thổ Nhĩ Kỳ, Thụy Điển, Tiếng Indonesia, Tiếng Ý, Trung, Trung (Phồn thể), Turkmen, Tây Ban Nha, Ukraina, Urdu, Uyghur, Uzbek, Việt, Xứ Wales, Yiddish, Yoruba, Zulu, Đan Mạch, Đức, Ả Rập, dịch ngôn ngữ.

Copyright ©2025 I Love Translation. All reserved.

E-mail: