3.2.3.4 Impact of Vessel Wall
There are two types of errors that can be associated with the placement of a flowmeter cuff on blood
vessels. The first source of these errors directly relates to the conductivity of the blood vessel wall, and
the second error can be attributed to the serous fluid between the blood vessel and the flowmeter cuff.17
Consideration has to be given to error arising from the conductivity of blood vessels and relates to
two main factors assuming that conductivity of blood is larger than 10–7 (ohm meter)-1. These two factors
are: (1) ratio of inside to outside diameter of the vessels and (2) the conductivity of blood to blood vessel
wall which has been determined to be of the order of 4.0 (i.e., blood is four times more conductive than
the vessel wall).18 Based on the fact that the typical conductivity ratio of blood to blood vessel wall is a
constant, then the error arising from this can specifically be ascribed to the ratio of inside and outside
diameter of the vessel in question where flow is being measured. The relationship between reduction in
signal (i.e., voltage as described in Equation 3.11) and ratio of the inside and outside diameter of the
blood vessel is relatively linear. For example, in vessels where the ratio of inside to outside diameter is
0.7, 0.8, and 0.9, the error is approximately 11, 9, and 5%, respectively.17 These are values by which the
voltage will deviate from the theoretical value, the actual impact being that a lower voltage will be detected
by the recording electrodes, which is attributed to electrical shunting. It is apparent that as the vessel
wall becomes thinner and the ratio of inside diameter to outside diameter reaches unity, the error is
substantially reduced. Thus, typical errors here are of the order of –5 to –10% deviation from the actual
1140_book.fm Page 9 Wednesday, July 9, 2003 8:46 AM
© 2004 by CRC Press LLC
3-10 Biomedical Technology and Devices Handbook
true signal generated by flow. It should also be mentioned that the theoretical calculation for this error
is based on Equation 3.11, which assumes the mean velocity for flow in a given cylinder, and more
importantly the fact that flow is rotationally symmetrical.17 Clearly, blood flow in blood vessels is not
always rotationally symmetrical (see Section 3.2.3.5).
Error associated with serous fluid between the blood vessel and the detecting electrodes of the flowmeter can occur due to polarization and electrical shunting. Any tissue fluid or blood lying between the
vessel and the flowmeter cuff can provide an additional electric shunt, which will ultimately manifest
itself as a reduction in the voltage detected by the electrodes. For example, for a vessel where the ratio
of inside to outside diameter is 0.8, a conductive film of a thickness of between 2 and 4% of the outside
radius of the vessel will result in error greater than 10% but less than 20%. For a vessel in which the
ratio of inside diameter to outside is 0.9, a conductive film of thickness between 2 and 4% of the outside
radius of the vessels will result in an error of 5 to 10%.17 Again, as with the error associated with vessels
conductivity, with the vessel wall becoming thinner and the ratio of inside diameter to outside diameter
reaching unity, the actual error occurring as a result of electric shunting due to conductive fluids between
the electrode and vessel wall will decline.17 To effectively reduce error due to electric shunting, the
flowmeter should be cuffed tightly around the blood vessel. It is obvious that a typical thickness of film
of conductive fluid is hard to predict, and error attributed to this factor has to be estimated for individual
vessels and each flowmeter cuff.
Experimentally, it has been found that meticulous preparation of a blood vessel for in vivo recording
of blood flow in small blood vessels in animals such as rats and rabbits improves zero-line stability.19
Essentially, careful dissection and cleaning of blood vessels allows for a mechanically good electrode-towall contact. There has been marked improvement in stability in blood flow measurement after removal
of the adventitial tissue. In addition, removal of the adventitial layer may remove serous fluid and thus
reduce the thickness of the conductive fluid outside the vessel which would be in contact with the
recording electrodes of the probe. Moreover, nonstatic serous fluid can also cause problems with zeroline stability due to polarization of the charge around the recording electrode. Certainly, the removal of
adventitial tissue will allow for much better contact of the probe with the blood vessel, pulling the ratio
of inside vessel diameter to outside diameter closer to unity as well. As already mentioned, the film of
conductive fluid outside the vessel and in contact with electrodes provides for electric shunting, thus
reducing the actual voltage recorded by the flowmeter.17 However, a smaller ratio between the inside and
outside diameter of blood the vessel will diminish the impact of this source of error, as well as that
associated with conductivity of the vessel. A foreseen problem associated with stripping the adventitial
tissue is the occurrence of spasm. In fact, spasm may occur and has been observed both proximal and
distal to the prepared area.19 Thus, careful cleaning and dissection of the vessel on which the probe is to
be placed is quite important, perhaps critical for accurate and reproducible blood flow measurements.
Clearly, changes in vessel wall thickness displace the ratio between probe diameter and inner vessel
diameter. This physical factor appears to be of considerable importance when dealing with small probes and
vessels even with constant electrode-to-wall contact. This also emphasizes the critical importance of selection
of the probe diameter for accurate measurements of blood flow using an electromagnetic flowmeter.
0/5000
Từ: -
Sang: -
Kết quả (Việt) 1: [Sao chép]
Sao chép!
3.2.3.4 Impact of Vessel WallThere are two types of errors that can be associated with the placement of a flowmeter cuff on bloodvessels. The first source of these errors directly relates to the conductivity of the blood vessel wall, andthe second error can be attributed to the serous fluid between the blood vessel and the flowmeter cuff.17Consideration has to be given to error arising from the conductivity of blood vessels and relates totwo main factors assuming that conductivity of blood is larger than 10–7 (ohm meter)-1. These two factorsare: (1) ratio of inside to outside diameter of the vessels and (2) the conductivity of blood to blood vesselwall which has been determined to be of the order of 4.0 (i.e., blood is four times more conductive thanthe vessel wall).18 Based on the fact that the typical conductivity ratio of blood to blood vessel wall is aconstant, then the error arising from this can specifically be ascribed to the ratio of inside and outsidediameter of the vessel in question where flow is being measured. The relationship between reduction insignal (i.e., voltage as described in Equation 3.11) and ratio of the inside and outside diameter of theblood vessel is relatively linear. For example, in vessels where the ratio of inside to outside diameter is0.7, 0.8, and 0.9, the error is approximately 11, 9, and 5%, respectively.17 These are values by which thevoltage will deviate from the theoretical value, the actual impact being that a lower voltage will be detectedby the recording electrodes, which is attributed to electrical shunting. It is apparent that as the vesselwall becomes thinner and the ratio of inside diameter to outside diameter reaches unity, the error issubstantially reduced. Thus, typical errors here are of the order of –5 to –10% deviation from the actual1140_book.fm Page 9 Wednesday, July 9, 2003 8:46 AM© 2004 by CRC Press LLC3-10 Biomedical Technology and Devices Handbooktrue signal generated by flow. It should also be mentioned that the theoretical calculation for this erroris based on Equation 3.11, which assumes the mean velocity for flow in a given cylinder, and moreimportantly the fact that flow is rotationally symmetrical.17 Clearly, blood flow in blood vessels is notalways rotationally symmetrical (see Section 3.2.3.5).Error associated with serous fluid between the blood vessel and the detecting electrodes of the flowmeter can occur due to polarization and electrical shunting. Any tissue fluid or blood lying between thevessel and the flowmeter cuff can provide an additional electric shunt, which will ultimately manifestitself as a reduction in the voltage detected by the electrodes. For example, for a vessel where the ratioof inside to outside diameter is 0.8, a conductive film of a thickness of between 2 and 4% of the outsideradius of the vessel will result in error greater than 10% but less than 20%. For a vessel in which theratio of inside diameter to outside is 0.9, a conductive film of thickness between 2 and 4% of the outsideradius of the vessels will result in an error of 5 to 10%.17 Again, as with the error associated with vesselsconductivity, with the vessel wall becoming thinner and the ratio of inside diameter to outside diameterreaching unity, the actual error occurring as a result of electric shunting due to conductive fluids betweenthe electrode and vessel wall will decline.17 To effectively reduce error due to electric shunting, theflowmeter should be cuffed tightly around the blood vessel. It is obvious that a typical thickness of filmof conductive fluid is hard to predict, and error attributed to this factor has to be estimated for individualvessels and each flowmeter cuff.Experimentally, it has been found that meticulous preparation of a blood vessel for in vivo recordingof blood flow in small blood vessels in animals such as rats and rabbits improves zero-line stability.19Essentially, careful dissection and cleaning of blood vessels allows for a mechanically good electrode-towall contact. There has been marked improvement in stability in blood flow measurement after removalof the adventitial tissue. In addition, removal of the adventitial layer may remove serous fluid and thusreduce the thickness of the conductive fluid outside the vessel which would be in contact with therecording electrodes of the probe. Moreover, nonstatic serous fluid can also cause problems with zeroline stability due to polarization of the charge around the recording electrode. Certainly, the removal ofadventitial tissue will allow for much better contact of the probe with the blood vessel, pulling the ratioof inside vessel diameter to outside diameter closer to unity as well. As already mentioned, the film ofconductive fluid outside the vessel and in contact with electrodes provides for electric shunting, thusreducing the actual voltage recorded by the flowmeter.17 However, a smaller ratio between the inside andoutside diameter of blood the vessel will diminish the impact of this source of error, as well as thatassociated with conductivity of the vessel. A foreseen problem associated with stripping the adventitialtissue is the occurrence of spasm. In fact, spasm may occur and has been observed both proximal anddistal to the prepared area.19 Thus, careful cleaning and dissection of the vessel on which the probe is tobe placed is quite important, perhaps critical for accurate and reproducible blood flow measurements.Clearly, changes in vessel wall thickness displace the ratio between probe diameter and inner vesseldiameter. This physical factor appears to be of considerable importance when dealing with small probes andvessels even with constant electrode-to-wall contact. This also emphasizes the critical importance of selectionof the probe diameter for accurate measurements of blood flow using an electromagnetic flowmeter.
đang được dịch, vui lòng đợi..
Kết quả (Việt) 2:[Sao chép]
Sao chép!
3.2.3.4 Tác động của Tàu tường
Có hai loại lỗi có thể được kết hợp với vị trí của một cuff lưu lượng kế trên huyết
mạch. Các nguồn đầu tiên của các lỗi này trực tiếp liên quan đến tính dẫn điện của thành mạch máu, và
các lỗi thứ hai có thể được quy cho các huyết thanh giữa các mạch máu và lưu lượng kế cuff.17
xem xét đã được trao cho lỗi phát sinh từ tính dẫn điện của mạch máu và liên quan đến
hai yếu tố chính giả định rằng độ dẫn máu lớn hơn 10-7 (mét ohm) -1. Hai yếu tố này
là: (1) tỷ lệ bên trong có đường kính bên ngoài của tàu và (2) độ dẫn của máu đến mạch máu
tường đó đã được xác định là số thứ tự của 4.0 (tức là, máu là bốn lần nữa dẫn hơn
thành mạch) .18 Dựa trên thực tế rằng tỷ lệ dẫn điển hình của máu vào thành mạch máu là một
hằng số, sau đó các lỗi phát sinh từ này có thể đặc biệt được cho là do tỷ lệ của các bên trong và bên ngoài
đường kính của tàu trong câu hỏi, nơi dòng chảy đang được đo. Mối quan hệ giữa giảm
tín hiệu (tức là, điện áp như được mô tả trong phương trình 3.11) và tỷ lệ của các bên trong và bên ngoài đường kính của
mạch máu là tương đối tuyến tính. Ví dụ, trong các mạch đó tỷ lệ của các bên trong có đường kính ngoài là
0,7, 0,8, 0,9, lỗi này là khoảng 11, 9, và 5%, respectively.17 Đây là những giá trị mà các
điện áp sẽ đi chệch khỏi các giá trị lý thuyết, tác động thực tế là có một điện áp thấp hơn sẽ được phát hiện
bởi các điện cực ghi âm, mà là do shunting điện. Rõ ràng là như tàu
tường trở nên mỏng hơn và tỷ lệ đường kính bên trong để đạt đến sự thống nhất đường kính bên ngoài, lỗi được
giảm đáng kể. Do đó, các lỗi điển hình ở đây là số thứ tự của -5 đến -10% độ lệch từ thực tế
1140_book.fm Page 9 Thứ Tư 9 Tháng 7, 2003 8:46
© 2004 bởi CRC Press LLC
3-10 y sinh Công nghệ và Thiết bị Handbook
đúng tín hiệu được tạo ra bởi dòng chảy. Nó cũng cần phải nói rằng các tính toán lý thuyết cho lỗi này
được dựa trên phương trình 3.11, trong đó giả định vận tốc trung bình cho lưu lượng trong một xi lanh được đưa ra, và nhiều hơn nữa
quan trọng thực tế là dòng chảy luân phiên symmetrical.17 Rõ ràng, dòng máu chảy trong mạch máu là không
luôn luôn luân phiên đối xứng (xem Phần 3.2.3.5).
Lỗi liên quan với huyết thanh giữa các mạch máu và các điện cực phát hiện của lưu lượng kế có thể xảy ra do sự phân cực và shunting điện. Bất kỳ dịch mô hoặc máu nằm giữa các
tàu và cuff lưu lượng kế có thể cung cấp một shunt điện bổ sung, mà cuối cùng sẽ thể hiện
chính nó như là một sự giảm điện áp được phát hiện bởi các điện cực. Ví dụ, đối với một tàu mà tỷ lệ
của các bên trong có đường kính ngoài là 0,8, một bộ phim dẫn điện có độ dày từ 2 đến 4% của bên ngoài
bán kính của tàu sẽ dẫn đến lỗi lớn hơn 10% nhưng dưới 20%. Đối với một con tàu, trong đó
tỷ lệ đường kính bên trong đến bên ngoài là 0,9, một bộ phim dẫn điện có độ dày từ 2 đến 4% của bên ngoài
bán kính của các tàu sẽ dẫn đến một lỗi của 5-10% .17 Một lần nữa, như với các lỗi liên hệ với mạch
dẫn, với các thành mạch trở nên mỏng hơn và tỷ lệ đường kính bên trong đến bên ngoài đường kính
đạt sự thống nhất, các lỗi thực sự xảy ra như là kết quả của shunting điện do chất lỏng dẫn điện giữa
các điện cực và thành mạch sẽ decline.17 Để giảm hiệu quả lỗi do shunting điện,
đồng hồ lưu lượng nên được còng chặt quanh các mạch máu. Nó được rõ ràng rằng một độ dày điển hình của bộ phim
của chất lỏng dẫn điện là khó dự đoán, và lỗi là do yếu tố này đã được ước tính cho từng
tàu và mỗi cuff lưu lượng kế.
Thực nghiệm, nó đã được tìm thấy rằng sự chuẩn bị tỉ mỉ của một mạch máu tại Vivo ghi
lưu lượng máu trong các mạch máu nhỏ ở động vật như chuột và thỏ cải thiện zero-line stability.19
Về cơ bản, bóc tách cẩn thận và làm sạch các mạch máu cho phép cho một số liên lạc điện-towall cơ học tốt. Hiện đã có cải thiện đáng kể trong sự ổn định trong đo lường lưu lượng máu sau khi loại bỏ
các mô adventitial. Ngoài ra, loại bỏ các lớp adventitial có thể loại bỏ huyết thanh và do đó
làm giảm độ dày của chất lỏng dẫn điện bên ngoài tàu mà sẽ được tiếp xúc với các
điện cực âm của tàu thăm dò. Hơn nữa, huyết thanh không tĩnh cũng có thể gây ra vấn đề với sự ổn định zeroline do sự phân cực của điện tích xung quanh điện cực âm. Chắc chắn, việc loại bỏ các
mô adventitial sẽ cho phép liên lạc tốt hơn nhiều của tàu thăm dò với các mạch máu, kéo tỷ lệ
đường kính bên trong tàu với đường kính bên ngoài gần hơn đến sự thống nhất là tốt. Như đã đề cập, bộ phim của
chất lỏng dẫn điện bên ngoài tàu và tiếp xúc với các điện cung cấp cho shunting điện, do đó
làm giảm điện áp thực tế ghi lại bởi các flowmeter.17 Tuy nhiên, một tỷ lệ nhỏ hơn giữa các bên trong và
đường kính ngoài của máu tàu sẽ giảm tác động của nguồn này cũng bị lỗi, cũng như
kết hợp với dẫn của tàu. Một vấn đề dự kiến kết hợp với tước adventitial
mô là sự xuất hiện của sự co thắt. Trong thực tế, sự co thắt có thể xảy ra và đã được quan sát thấy cả gần và
xa để các area.19 chuẩn bị Vì vậy, làm sạch cẩn thận và bóc tách của tàu mà tàu thăm dò là để
được đặt là khá quan trọng, có lẽ quan trọng để đo lưu lượng máu chính xác và lặp lại .
Rõ ràng, những thay đổi về độ dày thành mạch thuyên tỷ lệ giữa đường kính đầu dò và tàu bên trong
đường kính. Yếu tố vật lý này dường như là có tầm quan trọng đáng kể khi đối phó với các đầu dò nhỏ và
tàu ngay cả với liên tục tiếp xúc điện cực-to-tường. Điều này cũng nhấn mạnh tầm quan trọng của sự lựa chọn
của các đường kính đầu dò cho phép đo chính xác của lưu lượng máu bằng cách sử dụng một đồng hồ lưu lượng điện từ.
đang được dịch, vui lòng đợi..
 
Các ngôn ngữ khác
Hỗ trợ công cụ dịch thuật: Albania, Amharic, Anh, Armenia, Azerbaijan, Ba Lan, Ba Tư, Bantu, Basque, Belarus, Bengal, Bosnia, Bulgaria, Bồ Đào Nha, Catalan, Cebuano, Chichewa, Corsi, Creole (Haiti), Croatia, Do Thái, Estonia, Filipino, Frisia, Gael Scotland, Galicia, George, Gujarat, Hausa, Hawaii, Hindi, Hmong, Hungary, Hy Lạp, Hà Lan, Hà Lan (Nam Phi), Hàn, Iceland, Igbo, Ireland, Java, Kannada, Kazakh, Khmer, Kinyarwanda, Klingon, Kurd, Kyrgyz, Latinh, Latvia, Litva, Luxembourg, Lào, Macedonia, Malagasy, Malayalam, Malta, Maori, Marathi, Myanmar, Mã Lai, Mông Cổ, Na Uy, Nepal, Nga, Nhật, Odia (Oriya), Pashto, Pháp, Phát hiện ngôn ngữ, Phần Lan, Punjab, Quốc tế ngữ, Rumani, Samoa, Serbia, Sesotho, Shona, Sindhi, Sinhala, Slovak, Slovenia, Somali, Sunda, Swahili, Séc, Tajik, Tamil, Tatar, Telugu, Thái, Thổ Nhĩ Kỳ, Thụy Điển, Tiếng Indonesia, Tiếng Ý, Trung, Trung (Phồn thể), Turkmen, Tây Ban Nha, Ukraina, Urdu, Uyghur, Uzbek, Việt, Xứ Wales, Yiddish, Yoruba, Zulu, Đan Mạch, Đức, Ả Rập, dịch ngôn ngữ.

Copyright ©2025 I Love Translation. All reserved.

E-mail: