Agilent 7890A
Gas Chromatograph








Operating Guide

















Agilent Technologies

Notices
© Agilent Technologies, Inc. 2007-2010
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla- tion into a foreign language) without prior agreement and written consent from Agi- lent Technologies, Inc. as governed by United States and international copyright laws.

Manual Part Number
G3430-90011

Edition
Third edition, June 2010
Second edition, September 2008 First edition, March 2007
Printed in USA or China
Agilent Technologies, Inc. 2850 Centerville Road
Wilmington, DE 19808-1610 USA
Agilent Technologies, Inc. 412 Ying Lun Road Waigoaqiao Freed Trade Zone Shanghai 200131 P.R.China



Warranty
The material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor- mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control.



Safety Notices



A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.



A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Contents


1 Introduction



Where to Find Information 8 Online User Documentation 8
Chromatography Using a GC 10
The Front View of the Agilent 7890A GC 11 The Back View of the Agilent 7890A GC 11 The Inlets 12
The GC Column and Oven 15 Capillary Flow Technology 16 Detectors 17
The Operating Panel 18 The display 18
Status lights 19
Beeping instrument 19
Blinking setpoint 20
The keypad 21


2 Operating Basics
Overview 24
Instrument Control 25
Correcting Problems 26 To Start Up the GC 27
To Shut Down the GC for Less Than a Week 28 To Shut Down the GC for More Than a Week 29
3 Running a Method or a Sequence from the Keypad
Loading, Storing, and Running Methods from the Keypad 32 To load a method 32
To store a method 32
To manually inject a sample with a syringe and start a run 32 To run a method to process a single ALS sample 32
To abort a method 33
Loading, Storing, and Running Sequences from the Keypad 34 To load a sequence 34
To store a sequence 34

To start running a sequence 34 To pause a running sequence 35 To resume a paused sequence 35 To stop a running sequence 35
To resume a stopped sequence 35 Aborting a sequence 35
To resume an aborted sequence 36

4 Keypad Operation
The Run Keys 38
The Service Mode Key 38 The GC Component Keys 39 The Status Key 40
The Info Key 41
The General Data Entry Keys 42 The Supporting Keys 43
Method Storage and Automation Keys 44
Keypad Functionality When the GC Is Controlled by an Agilent Data System 45
About GC Status 46 Status Board 46
Alert tones 46
Error conditions 47
Blinking setpoint 47
About Logs 48
Maintenance log 48

5 Operating the 7693A Automatic Sampler
Placing Vials in the 7693A ALS 50 Placing vials in the sample tray 51
Placing vials in the turret (no sample tray) 52
Placing vials for sandwich injections 52 To Park the Injector 54
To Install a Syringe on the 7693A ALS 55
To Remove the Syringe on the 7693A ALS 58 To Configure Solvent/Waste Vials 59
7693A ALS Prerun Checklist 60

To Move the 7693A ALS to the Other GC Inlet 61 Check your work 62
Impact on your data system 62
To Adapt the 7693A ALS for the COC Inlet 63

6 Operating the 7683 Automatic Sampler
7683 ALS Turret Types 66
To Change the 7683 ALS Turret 67
To Move the 7683 ALS to the Other GC Inlet 69 To Park the Injector 71
To Install a Syringe on the 7683 ALS 72
To Remove the Syringe on the 7683 ALS 74 To Configure Solvent/Waste Vials 75
To Load Vials in the 7683 ALS Turret 77
To Load Vials in the 7683 ALS Sampler Tray 77 7683 ALS Prerun Checklist 78
To Adapt the 7683 ALS for the COC Inlet 79
To Install a 200-µm Column With the COC Inlet 80
To Manually Inject Onto a 200-µm Column With the COC Inlet 80

7 About Methods, Sequences, and Data Analysis
What Is a Method? 82
What Is Saved in a Method? 82
What Happens When You Load a Method? 83 What Is a Sequence? 83
Automating Data Analysis, Method Development, and Sequence Development 83



Agilent 7890A Gas Chromatograph Operating Guide




1
Introduction

Where to Find Information 8 Chromatography Using a GC 10
The Front View of the Agilent 7890A GC 11 The Back View of the Agilent 7890A GC 11 The Inlets 12
The GC Column and Oven 15 Capillary Flow Technology 16 Detectors 17
The Operating Panel 18





This document provides an overview of the individual components that make up the Agilent 7890A Gas Chromatograph (GC).
































Agilent Technologies 7



Where to Find Information

In addition to this document, Agilent provides several learning products that document how to install, operate, maintain, and troubleshoot the Agilent 7890A GC.
Before operating your GC, be sure to read the safety and regulatory information included on the Agilent GC and GC/MS Hardware User Information & Utilities DVD. The most common safety hazards when working on the GC are:
• Burns caused by touching heated areas on or in the GC
• Release of pressurized gas containing hazardous chemical compounds caused by opening inlets
• Glass cuts or puncture wounds caused by sharp capillary column ends
• Use of hydrogen as a GC carrier gas

Online User Documentation
Now your Agilent instrument documentation is in one place, at your fingertips.



The Agilent GC and GC/MS Hardware User Information & Utilities DVD that ships with your instrument provides an extensive collection of online help, videos, and books for current Agilent gas chromatographs, mass selective detectors, and GC samplers. Included are localized versions of the information you need most, such as:
• Getting Familiar documentation
• Safety and Regulatory guide
• Site Preparation information
• Installation information





• Operating guides
• Maintenance information
• Troubleshooting details




Chromatography Using a GC

Chromatography is the separation of a mixture of compounds into individual components.

There are three major steps involved with separating and identifying components of a mixture using a GC. They are:
1 Injecting a sample into the GC. (This takes place at the inlet.)
2 Separating the sample into individual components. (This takes place inside the column in the oven.)
3 Detecting what compounds were in the sample. (This is done in the detector.)
During this process, status messages from the Agilent 7890A GC are displayed, and user changes to parameter settings can be made through the operating panel.







Inlet








Oven



Detector




Operating panel


Each part of this process is described in brief on the following pages of this document. Refer to the Advanced User Guide for more details.




The Front View of the Agilent 7890A GC




Detector cover Detectors
Inlets


Valves (not shown) Display



Keypad


Power switch Oven Latch



The Back View of the Agilent 7890A GC




Oven exhaust vent Inlet and detector vent


Gas supply connections





Electronic cable connections





Oven cooling inlet Power connection




The Inlets

Inlets are where samples are injected into the GC. The Agilent 7890A GC can have a maximum of two inlets, identified as Front Inlet and Back Inlet.
A complete selection of inlets—split/splitless [0–100 psi and 0–150 psi], multimode, purged packed, cool- on- column, programmed temperature vaporization, and volatiles interface—are available.
The type of inlet chosen is based on the type of analysis being done, the type of sample being analyzed, and the column being used.




Samples can be injected into the inlets by hand using a syringe, or an automatic sampling device (such as an Agilent Automatic Liquid Sampler or Agilent Headspace Sampler).




Automatic injectors
The optional Agilent 7693A automatic liquid sampler (ALS) with a sample tray and bar code reader automates liquid sample processing. The modular design allows the autoinjector to be easily moved from one inlet to another or from one GC to another. The modular design also permits easy inlet maintenance.
The Agilent 7890A GC can accommodate up to two autoinjectors, identified as Front Injector and Back Injector.


Autoinjector




Turret






Sample Tray

(Bar code reader / mixer / heater not visible)











Automatic gas sampling valves
The sampling valves are simple mechanical devices that introduce a sample of fixed size into the carrier gas stream. Valves are most frequently used to sample gases or liquids in constantly flowing streams.
The Agilent 7890A GC can accommodate up to two gas sampling valves, identified as Valve # 1 and Valve #2.





The valves are located inside the gas sampling valve box.














Gas Sampling Valve box




The GC Column and Oven

GC columns are located inside a temperature- controlled oven. Generally,
0/5000
Từ: -
Sang: -
Kết quả (Việt) 1: [Sao chép]
Sao chép!
• Operating guides• Maintenance informatioChromatography Using a GC Agilent 7890AGas ChromatographOperating GuideAgilent Technologies Notices© Agilent Technologies, Inc. 2007-2010No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla- tion into a foreign language) without prior agreement and written consent from Agi- lent Technologies, Inc. as governed by United States and international copyright laws.Manual Part NumberG3430-90011EditionThird edition, June 2010Second edition, September 2008 First edition, March 2007Printed in USA or ChinaAgilent Technologies, Inc. 2850 Centerville RoadWilmington, DE 19808-1610 USAAgilent Technologies, Inc. 412 Ying Lun Road Waigoaqiao Freed Trade Zone Shanghai 200131 P.R.China WarrantyThe material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor- mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control. Safety Notices A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Contents 1 Introduction Where to Find Information 8 Online User Documentation 8Chromatography Using a GC 10The Front View of the Agilent 7890A GC 11 The Back View of the Agilent 7890A GC 11 The Inlets 12The GC Column and Oven 15 Capillary Flow Technology 16 Detectors 17The Operating Panel 18 The display 18Status lights 19Beeping instrument 19Blinking setpoint 20The keypad 21 2 Operating BasicsOverview 24Instrument Control 25Correcting Problems 26 To Start Up the GC 27To Shut Down the GC for Less Than a Week 28 To Shut Down the GC for More Than a Week 293 Running a Method or a Sequence from the KeypadLoading, Storing, and Running Methods from the Keypad 32 To load a method 32To store a method 32To manually inject a sample with a syringe and start a run 32 To run a method to process a single ALS sample 32To abort a method 33Loading, Storing, and Running Sequences from the Keypad 34 To load a sequence 34To store a sequence 34 To start running a sequence 34 To pause a running sequence 35 To resume a paused sequence 35 To stop a running sequence 35To resume a stopped sequence 35 Aborting a sequence 35To resume an aborted sequence 364 Keypad OperationThe Run Keys 38The Service Mode Key 38 The GC Component Keys 39 The Status Key 40The Info Key 41The General Data Entry Keys 42 The Supporting Keys 43Method Storage and Automation Keys 44Keypad Functionality When the GC Is Controlled by an Agilent Data System 45About GC Status 46 Status Board 46
Alert tones 46
Error conditions 47
Blinking setpoint 47
About Logs 48
Maintenance log 48

5 Operating the 7693A Automatic Sampler
Placing Vials in the 7693A ALS 50 Placing vials in the sample tray 51
Placing vials in the turret (no sample tray) 52
Placing vials for sandwich injections 52 To Park the Injector 54
To Install a Syringe on the 7693A ALS 55
To Remove the Syringe on the 7693A ALS 58 To Configure Solvent/Waste Vials 59
7693A ALS Prerun Checklist 60

To Move the 7693A ALS to the Other GC Inlet 61 Check your work 62
Impact on your data system 62
To Adapt the 7693A ALS for the COC Inlet 63

6 Operating the 7683 Automatic Sampler
7683 ALS Turret Types 66
To Change the 7683 ALS Turret 67
To Move the 7683 ALS to the Other GC Inlet 69 To Park the Injector 71
To Install a Syringe on the 7683 ALS 72
To Remove the Syringe on the 7683 ALS 74 To Configure Solvent/Waste Vials 75
To Load Vials in the 7683 ALS Turret 77
To Load Vials in the 7683 ALS Sampler Tray 77 7683 ALS Prerun Checklist 78
To Adapt the 7683 ALS for the COC Inlet 79
To Install a 200-µm Column With the COC Inlet 80
To Manually Inject Onto a 200-µm Column With the COC Inlet 80

7 About Methods, Sequences, and Data Analysis
What Is a Method? 82
What Is Saved in a Method? 82
What Happens When You Load a Method? 83 What Is a Sequence? 83
Automating Data Analysis, Method Development, and Sequence Development 83



Agilent 7890A Gas Chromatograph Operating Guide




1
Introduction

Where to Find Information 8 Chromatography Using a GC 10
The Front View of the Agilent 7890A GC 11 The Back View of the Agilent 7890A GC 11 The Inlets 12
The GC Column and Oven 15 Capillary Flow Technology 16 Detectors 17
The Operating Panel 18





This document provides an overview of the individual components that make up the Agilent 7890A Gas Chromatograph (GC).
































Agilent Technologies 7



Where to Find Information

In addition to this document, Agilent provides several learning products that document how to install, operate, maintain, and troubleshoot the Agilent 7890A GC.
Before operating your GC, be sure to read the safety and regulatory information included on the Agilent GC and GC/MS Hardware User Information & Utilities DVD. The most common safety hazards when working on the GC are:
• Burns caused by touching heated areas on or in the GC
• Release of pressurized gas containing hazardous chemical compounds caused by opening inlets
• Glass cuts or puncture wounds caused by sharp capillary column ends
• Use of hydrogen as a GC carrier gas

Online User Documentation
Now your Agilent instrument documentation is in one place, at your fingertips.



The Agilent GC and GC/MS Hardware User Information & Utilities DVD that ships with your instrument provides an extensive collection of online help, videos, and books for current Agilent gas chromatographs, mass selective detectors, and GC samplers. Included are localized versions of the information you need most, such as:
• Getting Familiar documentation
• Safety and Regulatory guide
• Site Preparation information
• Installation information





• Operating guides
• Maintenance information
• Troubleshooting details




Chromatography Using a GC

Chromatography is the separation of a mixture of compounds into individual components.

There are three major steps involved with separating and identifying components of a mixture using a GC. They are:
1 Injecting a sample into the GC. (This takes place at the inlet.)
2 Separating the sample into individual components. (This takes place inside the column in the oven.)
3 Detecting what compounds were in the sample. (This is done in the detector.)
During this process, status messages from the Agilent 7890A GC are displayed, and user changes to parameter settings can be made through the operating panel.







Inlet








Oven



Detector




Operating panel


Each part of this process is described in brief on the following pages of this document. Refer to the Advanced User Guide for more details.




The Front View of the Agilent 7890A GC




Detector cover Detectors
Inlets


Valves (not shown) Display



Keypad


Power switch Oven Latch



The Back View of the Agilent 7890A GC




Oven exhaust vent Inlet and detector vent


Gas supply connections





Electronic cable connections





Oven cooling inlet Power connection




The Inlets

Inlets are where samples are injected into the GC. The Agilent 7890A GC can have a maximum of two inlets, identified as Front Inlet and Back Inlet.
A complete selection of inlets—split/splitless [0–100 psi and 0–150 psi], multimode, purged packed, cool- on- column, programmed temperature vaporization, and volatiles interface—are available.
The type of inlet chosen is based on the type of analysis being done, the type of sample being analyzed, and the column being used.




Samples can be injected into the inlets by hand using a syringe, or an automatic sampling device (such as an Agilent Automatic Liquid Sampler or Agilent Headspace Sampler).




Automatic injectors
The optional Agilent 7693A automatic liquid sampler (ALS) with a sample tray and bar code reader automates liquid sample processing. The modular design allows the autoinjector to be easily moved from one inlet to another or from one GC to another. The modular design also permits easy inlet maintenance.
The Agilent 7890A GC can accommodate up to two autoinjectors, identified as Front Injector and Back Injector.


Autoinjector




Turret






Sample Tray

(Bar code reader / mixer / heater not visible)











Automatic gas sampling valves
The sampling valves are simple mechanical devices that introduce a sample of fixed size into the carrier gas stream. Valves are most frequently used to sample gases or liquids in constantly flowing streams.
The Agilent 7890A GC can accommodate up to two gas sampling valves, identified as Valve # 1 and Valve #2.





The valves are located inside the gas sampling valve box.














Gas Sampling Valve box




The GC Column and Oven

GC columns are located inside a temperature- controlled oven. Generally,
đang được dịch, vui lòng đợi..
Kết quả (Việt) 2:[Sao chép]
Sao chép!
• hướng dẫn điều hành
• Bảo trì informatioChromatography Sử dụng một GC Agilent 7890A máy sắc ký Hướng dẫn điều hành Agilent Technologies Notices © Agilent Technologies, Inc. 2007-2010 Không có một phần của hướng dẫn này có thể được sao chép dưới mọi hình thức hoặc bằng bất kỳ phương tiện (bao gồm lưu trữ điện tử và hồi hoặc transla -. tion sang tiếng nước ngoài) mà không có sự thoả thuận trước và được sự đồng ý bằng văn bản của Agi- vay Technologies, Inc là chi phối bởi Hoa Kỳ và luật bản quyền quốc tế Manual Part Number G3430-90011 bản ấn bản thứ ba, tháng 6 năm 2010 Thứ hai phiên bản, September 2008 phiên bản đầu tiên , tháng 3 năm 2007 In tại Mỹ hoặc Trung Quốc Agilent Technologies, Inc 2850 Centerville Road Wilmington, DE 19.808-1.610 USA Agilent Technologies, Inc. 412 Ying Lun Road Khu thương mại Waigoaqiao Freed Thượng Hải 200.131 PRChina Bảo hành Tư liệu có trong phát văn kiện này được cung cấp "như là," và là ject phụ để được thay đổi, mà không cần thông báo, trong các phiên bản trong tương lai. Hơn nữa, đến mức imum max- luật pháp cho phép, Agilent nhận bảo đảm, thể hiện rõ ràng hay ngụ ý, đối với hướng dẫn này và bất kỳ thông tin ở đây, bao gồm nhưng không giới hạn ở các bảo đảm ngụ ý khả năng bán được và phù hợp cho một biệt mục đích cá biệt. Agilent không chịu trách nhiệm về các lỗi hoặc các thiệt hại ngẫu nhiên hoặc do hậu quả trong kết nối với các trang trí nội thất, sử dụng, hoặc mance quả thực hiện của tài liệu này hoặc của bất kỳ thông tin trong tài liệu này. Nên Agilent và người dùng phải có văn bản thỏa thuận riêng biệt với các điều khoản bảo hành bao gồm các tài liệu trong tài liệu này mà xung đột với các điều khoản, điều kiện bảo hành trong các thỏa thuận arate sep- phải kiểm soát. An toàn Notices Thông báo CAUTION biểu thị một mối nguy hiểm. Nó kêu gọi sự chú ý đến một quy trình điều hành, thực tập, hoặc như thế, nếu không được thực hiện hoặc tham gia một cách chính xác, có thể dẫn đến thiệt hại cho các sản phẩm hoặc mất dữ liệu quan trọng. Không tiến xa hơn một thông báo CAUTION cho đến khi điều kiện chỉ được hiểu đầy đủ và đáp ứng. Một thông báo WARNING biểu thị một mối nguy hiểm. Nó kêu gọi sự chú ý đến một quy trình điều hành, thực tập, hoặc như thế, nếu không được thực hiện hoặc tham gia một cách chính xác, có thể dẫn đến thương tật hay tử vong. Không tiến xa hơn một thông báo WARNING cho đến khi các điều kiện đã định được hiểu đầy đủ và đáp ứng. Nội dung 1 Giới thiệu Nơi để Tìm thông tin 8 Online Documentation tài 8 Chromatography Sử dụng một GC 10 The Front View của Agilent 7890A GC 11 The Back Xem của Agilent 7890A GC 11 cửa hút gió 12 Các GC Cột và oven 15 mao mạch dòng chảy công nghệ 16 Detectors 17 Các điều hành Bảng điều chỉnh 18 Màn hình hiển thị 18 đèn Status 19 Beeping cụ 19 Blinking setpoint 20 Bàn phím 21 2 vấn đề cơ bản điều hành Tổng quan 24 Instrument kiểm soát 25 Chỉnh vấn đề 26 Để Start Up GC 27 Để Shut Down GC cho ít hơn một tuần 28 Để Shut Down GC cho More Than một Tuần 29 3 Chạy một phương pháp hoặc một trình tự từ bàn phím Đang tải, lưu trữ, và chạy các phương pháp từ bàn phím 32 Để tải một phương pháp 32 Để lưu trữ một phương pháp 32 Để bơm tay một mẫu với một ống tiêm và bắt đầu chạy 32 Để chạy một phương pháp để xử lý một ALS mẫu đơn 32 Để hủy bỏ một phương pháp 33 Đang tải, lưu trữ, và chạy Sequences từ bàn phím 34 Để tải một chuỗi 34 Để lưu trữ một chuỗi 34 Để bắt đầu chạy một chuỗi 34 Để tạm dừng một trình tự chạy 35 Để tiếp tục một chuỗi dừng 35 Để ngăn chặn một trình tự chạy 35 Để tiếp tục một chuỗi dừng 35 hủy bỏ một chuỗi 35 Để tiếp tục một chuỗi aborted 36 4 Bàn phím Operation Các Run phím 38 Các chế độ Service Key 38 GC Component phím 39 Status Key 40 Info Key 41 Tổng nhập dữ liệu phím 42 phím hỗ trợ 43 Phương pháp lưu trữ và tự động hóa phím 44 phím chức năng Khi GC Được điều khiển bởi một hệ thống Agilent dữ liệu 45 Về GC Status 46 Status Board 46 Alert tông 46 điều kiện Lỗi 47 Blinking setpoint 47 về Logs 48 Bảo trì đăng nhập 48 5 Vận hành 7693A Tự động Sampler Đặt lọ trong 7693A ALS 50 lọ Đặt trong khay mẫu 51 lọ Đặt trên tháp pháo (không có mẫu khay) 52 lọ Đặt tiêm bánh sandwich 52 Để Đậu các Injector 54 để cài đặt một ống tiêm trên 7693A ALS 55 để Hủy bỏ các Syringe trên 7693A ALS 58 Để cấu hình Solvent / Xử lý chất thải lọ 59 7693A ALS Prerun Checklist 60 để di chuyển các ALS 7693A để GC khác Inlet 61 Kiểm tra công tác 62 của bạn tác động vào hệ thống dữ liệu của bạn 62 Để thích ứng với các 7693A ALS cho COC Inlet 63 6 Vận hành 7683 tự động Sampler 7683 ALS Turret loại 66 Để thay đổi 7683 ALS Turret 67 Để Di chuyển 7683 ALS đến Khác GC Inlet 69 Để Đậu các Injector 71 để cài đặt một ống tiêm trên 7683 ALS 72 để Hủy bỏ các Syringe trên 7683 ALS 74 Để cấu hình Solvent / Xử lý chất thải lọ 75 Để tải lọ trong 7683 ALS Turret 77 Để tải lọ trong 7683 ALS Sampler Tray 77 7683 ALS Prerun Checklist 78 Để thích ứng với sự 7683 ALS cho COC Inlet 79 để cài đặt một cột 200 mm Với COC Inlet 80 Để Tiêm thủ Onto một cột 200 mm Với COC Inlet 80 7 Về phương pháp, Trình tự, và phân tích dữ liệu một phương pháp là gì? 82 là gì lưu trong một phương pháp? 82 Điều gì xảy ra khi bạn tải một phương pháp? 83, một trình tự là gì? 83 Tự động hoá phân tích dữ liệu, phát triển phương pháp, và phát triển chuỗi 83 Agilent 7890A máy sắc ký điều hành Hướng dẫn 1 Giới thiệu Nơi để Tìm thông tin 8 Chromatography Sử dụng một GC 10 The Front View của Agilent 7890A GC 11 The Back Xem của Agilent 7890A GC 11 cửa hút gió 12 Các Cột GC và Công nghệ Oven 15 mao mạch chảy 16 Detectors 17 Các điều hành Bảng điều chỉnh 18 Tài liệu này cung cấp một cái nhìn tổng quan về các thành phần riêng lẻ tạo nên Agilent 7890A sắc ký khí (GC). Agilent Technologies 7 ở đâu để Tìm thông tin Ngoài tài liệu này , Agilent cung cấp một số sản phẩm học tập, tài liệu về cách cài đặt, vận hành, bảo trì và khắc phục sự cố các Agilent 7890A GC. Trước khi hành GC của bạn, hãy chắc chắn để đọc các an toàn và thông tin pháp lý bao gồm trên Agilent GC và Thông tin GC / MS Phần cứng User & Utilities DVD. Các mối nguy an toàn phổ biến nhất khi làm việc trên GC là: • Burns gây ra bằng cách chạm vào các khu vực nóng trên hoặc trong GC • Phát hành khí nén có chứa hợp chất hóa học độc hại gây ra bằng cách mở cửa hút gió • cắt thuỷ tinh hoặc các vết thương đâm thủng gây ra bởi cột mao sắc nét kết thúc • Sử dụng hydro như một tàu sân bay GC khí trực tuyến Tài liệu tài Bây giờ tài liệu hướng dẫn cụ Agilent của bạn đang ở một nơi, ở trong tầm tay của bạn. Thông tin & Utilities Agilent GC và GC / MS Hardware tài DVD mà tàu với nhạc cụ của bạn cung cấp một bộ sưu tập các trợ giúp trực tuyến , video và sách cho chromatographs hiện Agilent khí, máy phát hiện có chọn lọc hàng loạt, và lấy mẫu GC. Bao gồm các phiên bản địa hoá của các thông tin mà bạn cần nhất, chẳng hạn như: • Bắt tài liệu quen thuộc • An toàn và tiết hướng dẫn • Sơ Chuẩn bị thông tin thông tin • Lắp đặt • Hướng dẫn điều hành • Bảo trì thông tin • chi tiết Troubleshooting sắc ký dùng một GC Sắc ký là việc tách một hỗn hợp các hợp chất thành các thành phần riêng lẻ. Có ba bước chính liên quan với tách và xác định thành phần của một hỗn hợp bằng cách sử dụng một GC. Đó là: 1 tiêm một mẫu vào GC. (Điều này diễn ra ở đầu vào.) 2 Tách các mẫu thành các thành phần riêng lẻ. (Điều này diễn ra bên trong các cột trong lò nướng.) 3 Phát hiện những hợp chất có trong mẫu. (Điều này được thực hiện trong các máy dò.) Trong quá trình này, các thông điệp trạng thái từ các Agilent 7890A GC được hiển thị, và người dùng thay đổi các thiết lập tham số có thể được thực hiện thông qua bảng điều hành. Inlet Oven Detector bảng Hoạt động Mỗi một phần của quá trình này được mô tả trong ngắn gọn trên trang sau của tài liệu này. Hãy tham khảo Hướng dẫn sử dụng chi tiết để biết thêm chi tiết. The Front View của Agilent 7890A GC Detector che Detectors cửa hút gió Van (không hiển thị) Hiển thị bàn phím điện chuyển Oven Latch The Back Xem của Agilent 7890A GC xả Oven trút Inlet và phát hiện lỗ thông cấp khí đốt kết nối điện tử, kết nối cáp Oven làm mát kết nối điện đầu vào cửa hút gió vịnh nhỏ là nơi các mẫu được tiêm vào GC. Agilent 7890A GC có thể có tối đa là hai cửa, xác định là Mặt trận Inlet và Back Inlet. Một lựa chọn hoàn hảo của cửa hút gió-split / splitless [0-100 psi và 0-150 psi], đa, được tẩy đóng gói, mát-on- cột, lập trình bay hơi nhiệt độ, và các chất bay hơi giao diện có sẵn. Các loại đầu vào được lựa chọn là dựa trên phương pháp phân tích được thực hiện, các loại mẫu được phân tích, và các cột được sử dụng. Các mẫu có thể được tiêm vào các cửa hút gió bằng tay sử dụng một ống tiêm, hoặc một thiết bị lấy mẫu tự động (chẳng hạn như một Agilent Automatic lỏng Sampler hoặc Agilent Headspace Sampler). kim phun tự động Các tùy chọn Agilent 7693A sampler chất lỏng tự động (ALS) với một khay mẫu và mã vạch đọc tự động hóa xử lý mẫu chất lỏng. Việc thiết kế mô-đun cho phép các autoinjector để dễ dàng di chuyển từ một đầu vào khác hoặc từ một GC khác. Việc thiết kế mô-đun cũng cho phép bảo dưỡng đầu vào dễ dàng. Các Agilent 7890A GC có thể chứa đến hai autoinjectors, xác định là Mặt trận Injector và Back Injector. Autoinjector Turret Tray Sample (Bar đọc mã / mixer / nóng không nhìn thấy được) lấy mẫu tự động bằng khí Van các loại van lấy mẫu là các thiết bị cơ khí đơn giản mà giới thiệu một mẫu có kích thước cố định vào trong dòng khí mang. Van được thường xuyên nhất được sử dụng để lấy mẫu khí hoặc chất lỏng trong các dòng suối không ngừng chảy. Các Agilent 7890A GC có thể chứa đến hai van lấy mẫu khí, xác định là Van # 1 và van 2 #. Các van được đặt bên trong hộp van lấy mẫu khí. Khí lấy mẫu Van hộp Các GC Cột và Oven cột GC được đặt bên trong một lò temperature- kiểm soát. Nói chung,

















































































































































































































































































































































































































đang được dịch, vui lòng đợi..
 
Các ngôn ngữ khác
Hỗ trợ công cụ dịch thuật: Albania, Amharic, Anh, Armenia, Azerbaijan, Ba Lan, Ba Tư, Bantu, Basque, Belarus, Bengal, Bosnia, Bulgaria, Bồ Đào Nha, Catalan, Cebuano, Chichewa, Corsi, Creole (Haiti), Croatia, Do Thái, Estonia, Filipino, Frisia, Gael Scotland, Galicia, George, Gujarat, Hausa, Hawaii, Hindi, Hmong, Hungary, Hy Lạp, Hà Lan, Hà Lan (Nam Phi), Hàn, Iceland, Igbo, Ireland, Java, Kannada, Kazakh, Khmer, Kinyarwanda, Klingon, Kurd, Kyrgyz, Latinh, Latvia, Litva, Luxembourg, Lào, Macedonia, Malagasy, Malayalam, Malta, Maori, Marathi, Myanmar, Mã Lai, Mông Cổ, Na Uy, Nepal, Nga, Nhật, Odia (Oriya), Pashto, Pháp, Phát hiện ngôn ngữ, Phần Lan, Punjab, Quốc tế ngữ, Rumani, Samoa, Serbia, Sesotho, Shona, Sindhi, Sinhala, Slovak, Slovenia, Somali, Sunda, Swahili, Séc, Tajik, Tamil, Tatar, Telugu, Thái, Thổ Nhĩ Kỳ, Thụy Điển, Tiếng Indonesia, Tiếng Ý, Trung, Trung (Phồn thể), Turkmen, Tây Ban Nha, Ukraina, Urdu, Uyghur, Uzbek, Việt, Xứ Wales, Yiddish, Yoruba, Zulu, Đan Mạch, Đức, Ả Rập, dịch ngôn ngữ.

Copyright ©2025 I Love Translation. All reserved.

E-mail: